@article{oai:miyazaki-u.repo.nii.ac.jp:00002772, author = {青木, 秀幸 and Aoki, Hideyuki and 太良尾, 浩生 and Tarao, Hiroo and 林, 則行 and Hayashi, Noriyuki and 青木, 秀幸 and Aoki, Hideyuki and 太良尾, 浩生 and Tarao, Hiroo}, journal = {宮崎大学工学部紀要, Memoirs of Faculty of Engineering, University of Miyazaki}, month = {Jul}, note = {It is necessary to assume a case in which human with a medical device made of metal (cardiac pacemaker, intramedullary nail etc.), experience electrical shock. Modelling of a medical device inside a model of human body and calculating a contact current inside the model one strongly demanded. In this study, we aimed to establish a technique to evaluate the precision of the calculation model including the metal body in human body. As this method, we selected a method for measuring the induced electric field inside the human model simulating the contact current. Constructing a system for the measurement and preparing two kinds of human models, we measured induced electric field inside the model. We confirmed the validity of this measurement system in comparison with the calculation result by the modified SPFD method result. As the result, it was found that our proposed system could measure the induced electric field with high accuracy.}, pages = {43--48}, title = {接触電流を模擬した生体モデル内における誘導電界の計測}, volume = {43}, year = {2014}, yomi = {アオキ, ヒデユキ and タラオ, ヒロオ and ハヤシ, ノリユキ and アオキ, ヒデユキ and タラオ, ヒロオ} }